GONOTEC GmbH

 

Advantage:

  • automatic calibration by switch
  • no ice formation in the lower cooling system: reliable measurements even for long working periods
  • measuring time approx. one minute: valuable in case of serial measurements
  • measurements and determination microprocessor controlled
  • minimum sample volume: 50µl (15µl-version optional available)
  • disposable plastic measuring vessels save time and costs
  • automatic error detection and display
  • optional measurement documentation by integrated matrix printer: date, time, sample no. and measuring result are printed

Technical Facts:

Standard Set

 

Duration of measurement

approx. one minute

Reproducibility

<±0.5% or <±2 digit for 50µl
<±1.0% or <±4 digit for 30µl

Display

4,5 digits

Measuring range

up to approx. 3000 mOsmol / kg

Resolution

1 mOsmol / kg over the entire measuring range

Cristallization process

by automatic injection of the sample with ice crystals

Cooling system

by means of two separate peltier cooling systems with electronical regulation of the lower cooling system (deviation <±0.1ºC)

Ambient temperature

10ºC to 35ºC

 

 

 

 

 

 

Options:

  • Option D: integrated matrix printer for automatic registration of the measurement result, date, time and sample no.
  • Option A: analog output for continuous recording of the temperature by a potentiometric recorder.
  • Option RS: digital output RS 232 C
  • Option M: sample volume of 15µl with reproducibility of <±2.0% or <±8 digits

Measurements Method

The total osmolality of aqueous solutions is determined by comparative measurements of the freezing points of pure water and of solutions. Whereas water has a freezing point of 0 °C, a solution with saline concentration of 1Osmol/kg has a freezing point of -1.858 °C.

Applications 

Routine measurements in the medical field as well as measurements in industry and research. The OSMOMAT 030 determines the total osmolality of aqueous solutions. The instruments require very small sample volumes and can thus be applied for extreme measuring tasks and serial measurements.

THE OSMOMAT 030 CAN BE USED IN:

  • General medicine
  • Routine and research
  • Forensic medicine
  • Electron microscopy
  • Physiology
  • Clinical Laboratories
  • Intensive care laboratories
  • Pediatrics
  • Gynaecology
  • Invitro fertilization (IVF)
  • Urology
  • Nephrology
  • Hemodialysis/Hemofiltration
  • Veterinary medicine
  • Botany
  • Pharmacy
  • Dispensari

For more information refer to site www.gonotec.com